Re: Comedi with SC-2345 and SCC-TC0x modules

> Thanks for responding.  I have already done everything you suggested.
> The PCI-602x are 12-bit boards yet when I read the binary values, I get
> values greater than 2048 and less than -2048 !?!?  So far, it would seem
> that comedi is the problem since it is reporting values beyond what the
> board should output.  I am using the suite of VIs from the NI website
> that use CINs (Comedi60waveformVIs.tgz).  If you can think of any other
> things I can try, please let me know.

Have you tried running the commands in the demo test suite from comedilib?  I 
use a large number of PCI-6023s and 6024s here, and have never had a problem.  
The analog inputs should give results from 0 to 4095.  My guess is that it's 
just a signed->unsigned number problem.

Caleb

Received on 2003-09-24Z12:50:37