- From: Noah Watkins <nwatkins_at_ittc.ku.edu>
- Date: Wed, 06 Jul 2005 23:37:44 -0500
Hi. I have a Measurement Computing PCI-DAS6025 board, though the question may not be specific to the board. In the User's Guide provided by the company, a settling time of 25 microseconds was typical for reading two different channels consecutively and acquiring good data in the worst case. I am experiencing with the cb_pcidas64 driver a situation where delays much longer have to be introduced. For example with nine volts applied to channel 8, and nothing applied to channel 0: Reading from channel 8: ~9V Reading from channel 0 to reset Sleep for 50 microseconds Reading from channel 0: ~7V In fact, it isn't until i sleep for around 80 _milli_seconds that i get readings that are reasonable. Given the sleep time limitation of instruction INSN_WAIT, all of my sleeping is occurring at the user level. This seems completely unrealistic. Has anyone ever encountered this degree of settling requirement before? Where might be a place to start investigating. Thanks, Noah W
Received on 2005-07-07Z03:37:44