Re: newbie repetetive comedi read questions

On Thu, 21.07.05, Per Arlock <Per.Arlock_at_med.lu.se> wrote:

>    cmd.convert_src := TRIG_TIMER;
>    cmd.convert_arg := 0;

Using the internal pacer with an arg of 0 is invalid. You have to pass the
sample rate (in ns) as cmd.convert_arg. This must be at least 1, but
probably higher since the 1710 can not sample once per nanosecond.

Nils

-- 
Institut fuer Schweisstechnik und Fuegetechnik (ISF)      Pontstrasse 49
der RWTH Aachen                                           D-52062 Aachen

ISF - Welding and Joining Institute          fon:  +49 (0)241 / 80-96274
RWTH - Aachen University                     fax:  +49 (0)241 / 80-92170

Received on 2005-07-21Z13:27:24