- From: Frank Mori Hess <fmhess_at_speakeasy.net>
- Date: Thu, 2 Feb 2006 20:11:19 -0500
On Thursday 02 February 2006 08:48 am, you wrote: > Frank: > > Here's the command I used: > > comedi_calibrate -v --calibrate --results /dev/comedi0 2>&1 > > comcal.txt > > and the file is attached. I'd be glad to do anything else necessary to > help. > Thanks. Your calibration actually doesn't look too bad, it is just the fine postgain offset and fine gain caldacs which can't quite get where they want to be. I think it is due to them not getting reset before the corresponding coarse caldac is set. Would you try cvs now? Or, just apply this patch: http://www.comedi.org/cgi-bin/viewcvs.cgi/comedilib/comedi_calibrate/ni.c?r1=1.138&r2=1.139 BTW, if I'm right, you should be able to work around this bug by using the --reset option. -- Frank
Attachments
- application/pgp-signature attachment: stored
Received on 2006-02-03Z01:11:19