- From: Apodaca, Bob <bob_at_phxlab.honeywell.com>
- Date: Fri, 03 Feb 2006 07:37:34 -0700
I applied the patch to my 0.72 version of comedilib and it worked. I can provide the output if it would be useful to anyone. Thanks for your support, Frank. Frank Mori Hess wrote: > On Thursday 02 February 2006 08:48 am, you wrote: > >>Frank: >> >>Here's the command I used: >> >>comedi_calibrate -v --calibrate --results /dev/comedi0 2>&1 > >>comcal.txt >> >>and the file is attached. I'd be glad to do anything else necessary to >>help. >> > > > Thanks. Your calibration actually doesn't look too bad, it is just the > fine postgain offset and fine gain caldacs which can't quite get where > they want to be. I think it is due to them not getting reset before the > corresponding coarse caldac is set. Would you try cvs now? Or, just > apply this patch: > > http://www.comedi.org/cgi-bin/viewcvs.cgi/comedilib/comedi_calibrate/ni.c?r1=1.138&r2=1.139 > > BTW, if I'm right, you should be able to work around this bug by using the > --reset option. >